User`s guide

7-73
Operating Concepts
TRL*/LRM* Calibration (ES Models Only)
Transmission magnitude uncertainty = E
X
+ E
T
S
21
+ E
S
S
11
S
21
+ E
L
S
22
S
21
where:
E
D
= effective directivity
E
R
= effective reflection tracking
E
S
= effective source match
E
L
= effective load match
E
X
= effective crosstalk
E
T
= effective transmission tracking
S
xx
= S-parameters of the device under test
The TRL Calibration Procedure
Requirements for TRL Standards
When building a set of TRL standards for a microstrip or fixture environment, the
requirements for each of these standard types must be satisfied.
Types Requirements
THRU
(Zero length) No loss. Characteristic impedance (Z
0
) need not be known.
•S
21
= S
12
= 1 0°
•S
11
= S
22
= 0
THRU
(Non-zero
length) Z
0
of the thru must be the same as the line. (If they are not the same,
the average impedance is used.)
Attenuation of the thru need not be known.
If the thru is used to set the reference plane, the insertion phase or
electrical length must be well-known and specified. If a non-zero length
thru is specified to have zero delay, the reference plane is established in
the middle of the thru, resulting in phase errors during measurement of
devices.
REFLECT Reflection coefficient Γ magnitude is optimally 1.0, but need not be
known.
Phase of Γ must known and specified to within ± 1/4 wavelength or
± 90°. During computation of the error model, the root choice in the
solution of a quadratic equation is based on the reflection data. An error
in definition would show up as a 180° error in the measured phase.
Γ must be identical on both ports.