Installation guide
Setting up a LAN Instrument Network  15 
Initialization Mode Selection 
The following initialization modes are selectable: 
NOTE 
  Compared to the G1369A LAN card, SW 7 and SW 8 must always be in OFF position on the 
G1369B LAN card, otherwise the selected modes are not working. 
Bootp 
When the initialization mode “Bootp” is selected, the card tries to 
download the parameters from a Bootp Server. The parameters 
obtained become the active parameters immediately. They are not 
stored to the non-volatile memory of the card. Therefore, the 
parameters are lost with the next power cycle of the card. 
Bootp & Store 
When “Bootp & Store” is selected, the parameters obtained from a 
Bootp Server become the active parameters immediately. In 
addition, they are stored to the non-volatile memory of the card. 
After a power cycle the parameters are still available. 
Using Stored 
When initialization mode “Using Stored” is selected, the parameters 
are taken from the non-volatile memory of the card. The TCP/IP 
connection will be established using these parameters. The 
parameters were configured previously by one of the methods 
described below. 
Using Default 
When “Using Default” is selected, the factory default parameters are 
taken instead. These parameters enable a TCP/IP connection to the 
LAN Interface without further configuration. 