Installation guide
Setting up a LAN Instrument Network  18 
Figure 8: Telnet information 
Information about the card 
Product id, firmware revision s, MAC address 
Initialization mode is Bootp. The connected PC/Bootserver is 
134.40.30.184 
Active TCP/IP settings 
Stored TCP/IP settings in non-volatile memory (not visible if equal to 
active TCP/IP settings) 
Connected to PC with controller software (e.g. Agilent ChemStation), 
here not connected 
5.  Change the IP address (in this example 134.40.24.230) and type 
“/” to list current settings. 
Figure 9: Change the IP address 
Enter ip 130.40.24.158 [Enter]. Changes the TCP/IP setting.  
Enter / [Enter]. It then displays these settings: 
Initialization mode is Bootp 
The connected PC/Bootserver is 134.40.30.184 
Active TCP/IP settings 
Stored TCP/IP settings in non-volatile memory 
Last user change IP 134.40.24.150 (not active yet, requires mode 
“Using Stored” and re-start) 
6.  When you have finished typing the configuration parameters, 
type: 
quit and press [Enter] to store the configuration parameters 
 or 
exit and press [Enter] to exit without storing parameters. 
If the Initialization Mode Switch is changed now to “Using 
Stored” mode, the instrument will take the stored settings 
when the module is re-booted. In the example above it would 
be 134.40.24.158 on QUIT and 134.40.24.160 on EXIT. 