Installation guide
Setting up a LAN Instrument Network  9 
 Instrument Firmware Requirements 
Table 1: Instrument Requirements 
Instrument Type  Model  Supported instrument 
firmware revision 
Method for setting IP 
addresses 
7890A  G3440A  A.01.10.3  Set at GC keypad 
7820A  G4350A  7820A.01.10.013.1.bin 
(E,C,J) 
7820A.01.10.013.Ru.bin 
Set using software 
keypad 
6890N GC  G1530N/ 
G1540N 
N.06.07 for 7693A 
And N.05.06 
LAN Assembly 
Firmware 04.7B3 
Set at GC keypad 
6890 Plus, 6890A  G1530A/ 
G1540A 
A.03.08 Chip Set  Determined by Jet 
Direct card used – see 
below. 
6850 Series GC 
SN≤US00003200 
G2630A 
A.03.07 for 7693A 
support and A.03.03 
Determined by Jet 
Direct card used – see 
below. 
6850 Series GC 
SN≥US10243001 
G2630A  A.06.02 for 7693A and 
A.05.04 
LAN Assembly 
Firmware 04.7B3 
Set at GC keypad 
35900E  35900E  E.01.02  Determined by Jet 
Direct card used – see 
below. 
Table 2: Jet Direct Card Requirements 
Product  Hardware 
Minimum 
Rev 
Support Statements 
G1369B  LAN Interface 
card 
B.01.01  10/100BaseT 
Supported with GC and 35900E 
Replaces the G1369A card mid 2010 
Set IP address with Telnet or BootP 
G1369A 
LAN Interface 
card 
C.03.00 
10/100BaseT 
Supported with GC and 35900E 
Set IP address with Telnet or BootP 
J4100A 
Jet Direct Card 
K.08.32 
10/100 Base T 
Set IP address with Telnet or BootP 
J2552A/B 
(obsolete) 
Jet Direct Card 
A.08.32 
10 Base T only 
Needs BootP Service (BootP must be running if 
the GC is cycled) 
LAN address is volatile (Upgrade to G1369B) 