Technical data
112 Understanding Your Agilent ChemStation
4 Integration
Manual Integration
In case the Manual Events checkbox of the Integration Events Table of a method is 
enabled, the manual events of the method are always applied when loading a 
data file using this method. If the data file contains additional manual events, 
they are applied after the events of the method. When the Manual Events 
checkbox is enabled, the user is never asked to save the events to the data file.
Conversion of Manual Integration Events Saved in a Method
In ChemStation revisions prior to B.04.01, manual integration events were 
stored in the method instead of the individual data file. In order to store the 
manual events in a data file specific way, it was most convenient to use the 
individual method DA.M of each specific data file.
The Integration menu in Data Analysis view provides the following items in order 
to either convert manual integration events in the method to events stored in 
the data file, or to continue to work with manual events stored in the method:
• Update Manual Events of Method: Save newly drawn manual events to the 
method.
• Apply Manual Events from Method: Apply the events currently saved in the 
method to the currently loaded data file.
• Remove Manual Events from Method: Delete the manual events from the 
method.