Technical data
126 Understanding Your Agilent ChemStation
5 Quantification
ISTD Calculation
RF
x
 is the Response Factor set in the Calibration Settings dialog box.
You can see from these formulae that the variations in the ISTD response 
are used to correct the quantification of the unknown component.
2 Use a calibrated peak. This ensures that the same response factor is used 
for the quantification of all peaks. The response factor of the selected 
compound and the uncalibrated peaks is corrected during all 
recalibrations. If the calibrated peak response factor changes, then the 
response factor for the unidentified peaks also changes by the same 
amount. If a Calibration Table is already set up, you can select a compound 
from the Using Compound combo box in the Calibration Settings dialog box.
The equations used to calculate the actual amount of an uncalibrated peak x 
are shown above.










