Technical data
Understanding Your Agilent ChemStation 137
Peak Identification
6
Peak Qualifiers
Signal Correlation
Signal correlation means that two peaks measured in different detector 
signals within a defined time window are assigned to the same compound. The 
signal correlation window can be controlled by the SignalCorrWin parameter in 
the QuantParm table of the _DaMethod register. Signal correlation is disabled 
when setting the signal correlation window to 0.0 minutes (see the Macro 
Programming Guide for more information). When signal correlation is off, 
peaks eluting at the same retention/migration time in different detector 
signals are treated as different compounds.
The default signal correlation window for LC, CE, CE/MS and LC/MS data is 
0.03 minutes and 0.0 minutes for GC data.
Qualifier Verification
If signal correlation is enabled, qualifier verification is active for all data file 
types by default. It can be disabled by setting the UseQualifiers flag in the 
Quantification Parameters table of the method (see the macro programming 
guide for more details). Qualifier verification is also disabled when signal 
correlation is switched off. 
Qualifier Ratio Calculation
When qualifiers verification is enabled for a compound, the ratio of the 
qualifier size and the main peak size is verified against the calibrated limits. 
The size may be height or area according to the calculation base setting in 
Specify Report.
The qualifier peaks can be calibrated in the same way as the target 
compounds. The user does not need to specify the expected qualifier ratio. The 
expected qualifier ratio is calculated automatically:
both measured at the retention time of the compound.
The QualTolerance parameter defines the acceptable range of the qualifier 
ratio, for example, ± 20%.










