6 Peak Identification
The Identification Process
Finding the Remaining Calibrated Peaks
The third pass identifies all remaining peaks listed in the calibration table. 
The non-reference peaks in the calibration table are matched to the remaining 
run peaks by using their RT window.
Each non-reference calibrated peak has its own retention/migration time in 
the calibration table. This is adjusted for the particular run based on the 
pre-identification of the time reference peaks. The retention/migration time 
window of the calibrated peak is adjusted based on the corrected 
retention/migration time of the calibrated peak.
If more than one peak is found in the same window, the peak with a 
retention/migration time which is closest to the expected retention/migration 
time and also meets the optional qualifier specifications is chosen.
Classification of Unidentified Peaks 
If there are remaining peaks, which are still not identified, they are classified 
as unknown. The ChemStation attempts to group the unknown peaks that 
belong to the same compound. If a peak has been detected in more than one 
signal, the peaks with the same retention/migration time in each signal are 
grouped to one compound.
Unknown peaks are reported if the corresponding selection has been made in 
the Calibration Settings dialog box.