Technical data
Understanding Your Agilent ChemStation 155
Calibration
7
Recalibration
Recalibration
What is Recalibration?
Recalibration is the process used when you want to update a level on a 
calibration curve. When you recalibrate you run another sample that contains 
the same calibration compounds as the original, and most important, the same 
amount of these compounds. When you run the calibration sample, you obtain 
updated response factors and retention/migration times. You may also choose 
to average the response factors over a number of calibration runs so that 
response factors are weighted equally.
Why Recalibrate?
Most calibrations have a limited lifetime, due to changes in chromatography. 
Recalibration is necessary to maintain the accuracy of the analysis. For 
example, assume you have created a calibration table for the compound 
caffeine which you use whenever you are required to quantify samples 
containing caffeine. At some point you will need to replace the column/ 
capillary. Although the column/capillary is replaced with exactly the same 
type, it will not behave in exactly the same way as the previous column/ 
capillary when you first created the calibration table for caffeine. Therefore, to 
ensure consistency you should recalibrate the levels in the calibration table 
before using the new column/capillary to analyze samples containing 
unknown amounts of caffeine. By doing this you are quantifying samples 
analyzed under the same system conditions.
Manual Recalibration
You can enter peak calibration information manually and normalize the 
calibration table using the Manual Setup option button in the New Calibration 
Table dialog box. Typically, a new calibration method is produced by running a 
calibration standard mixture, creating a calibration table, and entering the 
amounts of all calibrated peaks to obtain response factors. This approach is 
inefficient for some application, such as found in the petrochemical industry, 










