Technical data

180 Understanding Your Agilent ChemStation
8 Automation
Specifying Recalibrations
Specifying Recalibrations
The recalibration parameters for the sequence are entered directly into the
sequence table. These parameters define how the method is recalibrated
during a sequence.
Recalibration Parameters in the Sequence Table
The response factor and retention/migration times can be updated in several
ways. The calibration level, update response factor and update
retention/migration times are instructions used in the data analysis when
recalibrating the calibration table.
When Calibration is entered in the SampleType column of the sample table,
the following columns become active and can be edited:
Cal Level
Update RT
Update RF
Interval
The values that can be entered into each of these columns are shown in the
table.
Table 1 5 Recalibration Parameters in Sequence Table
CAL Level Update RT Update RF Interval
Calibration table Level
# (1-999)
No update No update Cyclic recalibration interval
# (1-999)
Average Average Blank
Replace Replace
Bracket
Delta%