Technical data
180 Understanding Your Agilent ChemStation
8 Automation
Specifying Recalibrations
Specifying Recalibrations
The recalibration parameters for the sequence are entered directly into the 
sequence table. These parameters define how the method is recalibrated 
during a sequence.
Recalibration Parameters in the Sequence Table 
The response factor and retention/migration times can be updated in several 
ways. The calibration level, update response factor and update 
retention/migration times are instructions used in the data analysis when 
recalibrating the calibration table. 
When Calibration is entered in the SampleType column of the sample table, 
the following columns become active and can be edited:
• Cal Level
• Update RT
• Update RF
• Interval
The values that can be entered into each of these columns are shown in the 
table.
Table 1 5 Recalibration Parameters in Sequence Table
CAL Level Update RT Update RF Interval
Calibration table Level 
# (1-999)
No update No update Cyclic recalibration interval 
# (1-999)
Average Average Blank
Replace Replace
Bracket
Delta%