Technical data
Understanding Your Agilent ChemStation 181
Automation
8
Specifying Recalibrations
The table shows the columns in the sequence table that contain the 
recalibration parameters and the values that can be entered.
No Update 
Does not change the response factor or retention/migration time.
Replace 
Replaces the previous retention/migration times and the response (areas or 
heights) with those from the current run only. The response is not changed for 
any peak that is not found in this recalibration run.
Average 
Averages the retention/migration times and responses (areas or heights) for 
each peak based on the original calibration run and all averaged recalibrations 
since then. If a peak is missing in one of the recalibrations, the average 
response of the peak will not suffer.
Bracket 
The samples are bracketed by pre-sample and post-sample calibrations. 
Evaluation is done when the last calibration sample of the closing bracket has 
been run. The existing calibration data is replaced by the result data of the 
calibration run of the opening bracket. The closing bracket calibrations are 
averaged to that calibration table.
Interval 
The Interval determines how often a calibration is done during a sequence. 
Calibration frequency corresponds to the number of sample injections that are 
done before the next set of calibration injections takes place. At the start of 
the analysis, a calibration is done and the results (response factors) are 
entered into the calibration table. These results are then used in subsequent 
quantitative calculations. After the specified number of injections have been 
done, another calibration is made and the results are entered into the 
calibration table overwriting the results of the previous calibration run.










