Technical data
190 Understanding Your Agilent ChemStation
8 Automation
Explicit and Cyclic Calibrations Together
Explicit and Cyclic Calibrations Together
This type of sequence comprises explicit and cyclic calibrations in the same 
sequence.
This feature allows you to recalibrate the method completely at the beginning 
of a sequence (explicit recalibration) and then update the calibration (cyclic 
recalibration) during the sequence.
• Two calibration lines for each calibration level in the Sequence table must 
be specified. One calibration line is for the explicit recalibration entry and 
the other for the cyclic recalibration entry.
• The sequence table must have entries for each calibration line and all cyclic 
recalibration vials must appear before both the explicit recalibration and 
sample entries.
Example 
The sequence table below illustrates a single-level calibrated method called 
SimpReg. It is truncated to simplify the example. 
There are two entries for the single calibration level.
Table 1 9 Sequence Table for SIMPREG
1 1 SimpReg 1 Calibration 1 Average Average 3
2 1 SimpReg 1 Calibration 1 Replace Replace
32SimpReg 1
43SimpReg 1
54SimpReg 1
65SimpReg 1
76SimpReg 1