Technical data

190 Understanding Your Agilent ChemStation
8 Automation
Explicit and Cyclic Calibrations Together
Explicit and Cyclic Calibrations Together
This type of sequence comprises explicit and cyclic calibrations in the same
sequence.
This feature allows you to recalibrate the method completely at the beginning
of a sequence (explicit recalibration) and then update the calibration (cyclic
recalibration) during the sequence.
Two calibration lines for each calibration level in the Sequence table must
be specified. One calibration line is for the explicit recalibration entry and
the other for the cyclic recalibration entry.
The sequence table must have entries for each calibration line and all cyclic
recalibration vials must appear before both the explicit recalibration and
sample entries.
Example
The sequence table below illustrates a single-level calibrated method called
SimpReg. It is truncated to simplify the example.
There are two entries for the single calibration level.
Table 1 9 Sequence Table for SIMPREG
1 1 SimpReg 1 Calibration 1 Average Average 3
2 1 SimpReg 1 Calibration 1 Replace Replace
32SimpReg 1
43SimpReg 1
54SimpReg 1
65SimpReg 1
76SimpReg 1