Technical data

Understanding Your Agilent ChemStation 193
Automation
8
Cyclic Calibration Sequences with Bracketing
Figure 41 Bracketing
Bracketed Sequence Operation
The first calibration vials are analyzed.
The sample vials are analyzed.
The next calibration vials are analyzed.
The calibration table is produced by replacing the existing response factors
by new ones and averaging the following calibration runs into a new
calibration table.
The sample vial data files are evaluated, and reports are generated.
The sequence returns to step 2 if there are more sample vials to be
analyzed.
Example
This section describes a sequence bracketing example comprising one method
called Brack.M. The Brack.M method is a two-level internal standard method
using cyclic calibration.