Technical data
Understanding Your Agilent ChemStation 205
Data Review, Reprocessing and Batch Review
9
Navigation Table in Data Analysis
is visible in the Status Bar; the system adds from data file in brackets (from 
data file) to indicate that the loaded method is the individual method for 
the data file.
2 Review your data using the sequence method: Use the option Sequence 
Method in the Preferences / Signal/Review Options to have the system load the 
sequence method corresponding to the current line of the Navigation Table. 
This method is loaded whenever a data file is loaded and is used for 
reviewing and generating the report. The method name is visible in the 
Status Bar; the system adds sequence in brackets (sequence) to indicate 
that the loaded method is the sequence method corresponding to the 
current line of the Navigation Table.
3 Review your data using a different method: If you want to use a different 
method for reviewing the data than the individual data analysis method 
(DA.M) stored with the data file or the sequence method, the option current 
method in Preferences / Signal/Review Option must be selected. In this case, 
the system uses the currently loaded method for reviewing and generating 
the report. The method name is visible in the Status Bar.
Sequence Reprocessing Using the Navigation Table
For reprocessing using the Navigation Table in Data Analysis, all necessary files 
are present in the sequence data container:
• sequence data files (*.d)
• all methods (*.m) files used during the sequence
• copy of the original sequence template (*.s)
NOTE
For LC, CE, LC/MS and CE/MS systems, the option Individual Method from Data File 
(DA.M) is selected by default.
For GC systems, the option Current Method is selected by default.
NOTE
Sequence data acquired with ChemStation revisions up to B.01.03 need to be reprocessed 
using the reprocess option in the Method and Run Control view. The same applies to data 
acquired in B.03.01 when Unique Folder Creation is switched off.
Sequence data acquired with ChemStation revisions B.02.01 and higher need to be 
reprocessed using the reprocessing toolset in the Data Analysis Navigation Table.