Technical data
206 Understanding Your Agilent ChemStation
9 Data Review, Reprocessing and Batch Review
Navigation Table in Data Analysis
• sequence-related batch (*.b) file
• sequence-related logbook (*.log) file
During reprocessing, the individual methods DA.M for the data files are 
updated as well as the batch file (*.b) file.
With the Data Analysis reprocessing functions, it is possible to modify the 
sequence template (*.s) in the data container in order to change the multiplier, 
dilution etc., or to use a different method for reprocessing. By default, the 
Data Analysis reprocessing sequence parameter parts of method to run is set to 
Reprocessing only, and the option Use Sequence Table Information is checked. 
These predefined default values enable you to change the parameters in the 
sequence template and run a reprocess without editing the  Data Analysis 
sequence parameters again.
If you have not explicitly changed the method in the sequence template, the 
system uses the sequence methods stored in the sequence data container to 
reprocess the sequence. These methods are the original methods used during 
data acquisition. Note that, even if the option load DA method from data file in 
Preferences / Signal/Review Options is selected, the system uses the sequence 
container method for reprocessing, not the DA.M of the individual data files.
If particular method parameters need to be changed (for example, specify to 
print to a *.xls file), the methods in the sequence container need to be 
modified and saved. This general change is then applied to all data files during 
reprocessing.
If you now want to use the updated sequence container method for further 
data acquisition, you need to copy this method from the sequence data 
container to one of the defined method paths. The new/updated method is 
then available in the ChemStation Explorer in the method view as a master 
method.