Technical data
Understanding Your Agilent ChemStation 211
Data Review, Reprocessing and Batch Review
9
Review Functions
Review Functions
Data files can be displayed in one of two ways:
• manually, by selecting a run to display from the table, 
• automatically, with a predefined interval between each data file. During 
automatic display, only those sample types displayed in the table are 
displayed; the runs are displayed in the order in which they appear in the 
table. The automatic review can be paused and later resumed, or stopped.
The standard functions provided by the ChemStation are available with batch 
review. This includes calibration, manual manipulation of chromatograms, for 
example by smoothing or manual integration. Any changes made to a data file 
can be marked and saved with the batch file. Chromatograms that have been 
reviewed are marked with an asterisk in the batch table. You can also discard 
changes made to either the current chromatogram only, or all changes to all 
chromatograms in the batch.
When a run is loaded, the selected processing options are performed; if the 
run has already been processed and the changes saved, the processed run is 
loaded. This is a faster process than loading the unprocessed run, because no 
processing needs to be done.
Calibration in Batch Review
Calibration in batch review works independently from the recalibration 
settings in the sequence table. The first step in batch calibration always 
replaces both response and retention time entries in the calibration table. For 
the following calibration standards, both response and retention time values 
are averaged.