Technical data
Understanding Your Agilent ChemStation 215
Using the ChemStation Reports
10
Reporting Results
Reporting Results
Two types of reports are available:
• an uncalibrated report which does not correct for detector response, and
• a calibrated report that shows results corrected for the difference in the 
detector response to various components of the sample.
Uncalibrated Reports 
The uncalibrated reports include the Area% and Height% reports. These reports 
are mainly used in the preparation of calibrated reports. They may be of value 
as a final report if the amounts of compound required to produce a unit area 
or height response for the compounds of interest are similar.
Calibrated Reports 
Calibrated reports correct for the difference in the detector response to the 
reported compounds. One or more calibration samples containing known 
amounts of the reported compounds must be run under the same conditions 
used for the unknown sample. The integration data from these calibration 
sample(s) are used to prepare a calibration table. This is a list of 
retention/migration times, amounts and responses which is used in the 
generation of the report. The calibrated reports are based on two calibration 
procedures called external standard and internal standard.
External Standard Report 
The ESTD report lists the results with the units of your choice or with each 
compound as a percentage of all the compounds present. The external 
standard procedure requires that the relative injected volume of both the 