Technical data
220 Understanding Your Agilent ChemStation
10 Using the ChemStation Reports
Report Styles
Report Styles 
The following report styles are available:
You choose to add a signal to any of the report styles by checking the 
appropriate box in the Specify Report dialog box.
• None  - No text will be reported. The chromatogram will be reported only if 
the Add Chromatogram Output option is selected.
• Short - Comprises quantitative text results of all integrated signals which 
are setup in the Signal Details dialog box (LC only) or the Signal dialog box 
(GC only). The peak width in the short report is calculated with the more 
complex formula used by the integrator: PW = 0.3(IPRight - IPLeft) + 
0.7(Area/Height) where IPRight and IPLeft are the inflection points.
• Detail - Comprises header, quantitative results, and calibration curves. The 
header is stored in a file called RPTHEAD.TXT in the method directory. You 
can change the header using a text editor to include a method specific text.
• Header + Short - Comprises file header and quantitative text results. The 
header is stored in a file called RPTHEAD.TXT in the method directory. You 
can change the header using a text editor to include a method specific text.
• GLP + Short - Comprises header, sample information, instrument conditions, 
logbook, signal, and quantitative results. The header is stored in a file called 
RPTHEAD.TXT in the method directory. You can change the header using a 
text editor to include a method specific text.
• GLP + Detail - Comprises header, sample information, instrument conditions, 
logbook, signal, quantitative results, and calibration curves. The header is 
stored in a file called RPTHEAD.TXT in the method directory. You can 
change the header using a text editor to include a method specific text.
• Full - Comprises header, sample information, instrument conditions, 
logbook, signals, and quantitative results. The header is stored in a file 
called RPTHEAD.TXT in the method directory. You can change the header 
using a text editor to include a method specific text.
• Performance - Produces a report according to the limits specified in the Edit 
Performance Limits dialog box of the System Suitability menu.