Technical data
226 Understanding Your Agilent ChemStation
10 Using the ChemStation Reports
Sequence Summary Reporting
Sequence Summary Reporting
Overview
The ChemStation can print a variety of standard reports for individual sample 
analyses. Sequence summary reporting is an additional way of reporting, that 
allows you to calculate and report parameters across a number of different 
analyses. It is useful, for example, to test the stability of an instrument or the 
robustness of a new method.
A sequence summary report may include:
• a title page,
• instrument configuration, including the revision numbers of the instrument 
and details of the analytical column/capillary being used,
• sequence table listings which describe what the automated sequence of 
analyses should have done,
• logbook descriptions of what the sequence actually did and any unexpected 
events which occurred during the sequence,
• method listings,
• individual reports on each sample,
• statistics on the analyses based on selected criteria—statistics are 
calculated only for calibrated compounds, and
• a table of contents with page number references to the detailed sections of 
the report.
Setting up a Sequence Summary Report
When setting up a sequence summary report, you can select any combination 
of the following nine categories by activating the appropriate check boxes and, 
where appropriate, selecting a report style from the template selections. Each 
template specifies the contents and layout of that particular section of the 
whole sequence summary report.
You can choose one of the following sequence summary report styles: