Technical data
Understanding Your Agilent ChemStation 233
Evaluating System Suitability
11
Sequence Summary Reporting
• peak width at half height, 
• peak symmetry,
• peak tailing, 
• capacity factor (k´), 
• plate numbers, 
• resolution between peaks, 
• selectivity relative to preceding peak, 
• skew, and
• excess. 
The mean value, the standard deviation, the relative standard deviation and 
the confidence interval are calculated. You can set limits for either standard 
deviation, the relative standard deviation or the confidence interval for each 
of these parameters. Should the values exceed your limits, the report is 
flagged to draw your attention to them.
The quality of the analytical data can be supported by keeping records of the 
actual conditions at the time the measurements were made. The 
ChemStation’s logbook records instrument conditions before and after a run. 
This information is stored with the data and reported with sample data. 
Instrument performance curves are recorded during the entire analysis as 
signals, and stored in the data file. If supported by the instrument these 
records, overlaid on the chromatogram, can be recalled on demand, for 
example, during an audit.
Baseline noise and drift can be measured automatically. A minimum 
detectable level can be calculated from peak height data for each calibrated 
compound in the method.
Finally, instrument configuration, instrument serial numbers, 
column/capillary identification, and your own comments can be included in 
each report printed.
Extended performance results are calculated only for compounds calibrated 
for in the method, ensuring characterization by retention/migration times and 
compound names.
A typical system performance test report contains the following performance 
results: 
• instrument details,










