Technical data
30 Understanding Your Agilent ChemStation
1 Agilent ChemStation Features
About the ChemStation Software
All reports have time stamps and traceable page numbering (page x of y 
pagination style). The user may select the level of detail in each report ranging 
from simple summary reports to complete system details (see the Reporting 
section above).
GLP Save register files, specified as part of the method configuration, save all 
the original data, including sample information, data analysis method, 
chromatographic/electropherographic signals, instrument conditions, 
integration and quantification results, report data and the run logbook in one 
checksum protected binary file. This is an uneditable binary format that 
ensures the originality of the results.The file includes a revisioning scheme 
that indicates if data has been reprocessed.
Control sample types may be defined in the sequence table and used to 
automatically check the instrument performance against quality control 
sample results when the instrument is running unattended. Results that are 
outside the user-specified acceptable range will stop the automatic execution 
of the instrument.










