Technical data
Understanding Your Agilent ChemStation 75
Integration
4
Overview
Allocating the Baseline
The integrator allocates the chromatographic/electropherographic baseline 
during the analysis at a frequency determined by the peak width value. When 
the integrator has sampled a certain number of data points, it resets the 
baseline from the initial baseline point to the current baseline point. The 
integrator resumes tracking the baseline over the next set of data points and 
resets the baseline again. This process continues until the integrator identifies 
the start of a peak
Figure 4 Baseline
At the start of the run, this baseline setting is used as the beginning baseline. 
If this is not set, the first data point is used. This baseline point is then 
periodically reset according to the following formula:
Areas are summed over a time T (expected peak width). This time can never be 
shorter than one data point. This continues as long as baseline condition 
exists. Slope and curvature are also taken. If both slope and curvature are less 
than the threshold, two summed areas are added together, and compared with 
the previous baseline. If the new value is less than the previous baseline, the 
new value immediately replaces the old one. If the new value is greater than 
the previous value, it is stored as a tentative new baseline value and is 
confirmed if one more value satisfies slope and curvature flatness criteria. 
This latter limitation is not in effect if negative peaks are allowed. During 
baseline, a check must also be made to examine fast rising solvents. They may 
be too fast for upslope detection. (By the time upslope is confirmed, solvent 
criterion may no longer be valid.) At first time through the first data point is 
baseline. It is replaced by the 2 T average if signal is on base. Baseline is then 
reset every T (see Figure 4 on page 75).
hiVgii^bZ
XjggZcied^ci
'I
'I
'I
'I
gZhZiWVhZa^cZ
:meZXiZYeZV`l^Yi]
9ViVed^ci










