User manual

28 User Manual
4 Instrument Connection
Analog Input (Signal) Connection
The signal must be within the range of -18 millivolts to 1 volt. A simple voltage
divider resistor network may be used to attenuate larger signals to this range.
Each input has three pins to which the high (+) and low (–) sides of the signal
and the shield are connected.
Use an instrument-specific cable (Table 6) whenever possible. Use a general
purpose cable in those cases where a custom cable is not available.
Appendix A lists pin assignments for all open cables.
1 Turn off the power to the instrument supplying the analog output signal.
2 For instrument-specific cablesConnect one end of the cable to the
instrument's analog output using the instrument manufacturer's
instructions.
For general purpose cablesUse the pin assignments shown in
Appendix A, “Cables, and the instrument manufacturer's instructions to
connect the two signal wires and ground to the instrument.
3 Connect the cable to the appropriate (A or B) analog input on the rear
panel of the 35900 (Figure 8).
4 If needed, repeat for the other channel.
CAUTION
For correct operation of the 35900, the input voltage must be in the range of -18 me to
+1 V. Voltages below –10 V or above +10 V may damage the instrument.