User`s guide
Power Tests Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 121
b Upper Transmission Band: λ
3
 ... λ
4
•TLS_λ
3 
= TLS_λ
0 
+ 0.5 - Attenuation Band
= TLS_λ
0 
+ 1 nm
•TLS_
λ
4 
= TLS_λ
0 
+ 0.5 x Upper Transmission Band
= TLS_λ
0 
+ 1 nm + 3 nm
8 Determine the maximum transmitted power value inside the 
transmission band.
a Record Spectrum.
b Using the marker, find the maximum transmitted power 
(max_SSE_power) within the Lower and Upper Transmission Bands. 
Change 
λ by using the RPG and note the maximum power value with-
in both the Lower and Upper Transmission Bands (this is a single val-
ue). Note this value in the Test Record. Check the associated 
wavelength from the Optical Specrum Analyzer, 
OSA@max_SSE_power, and note this value in the Test Record.
9 Set the marker of the OSA to OSA@max_SSE_power.
Change [
λ], the output wavelength of the TLS, so that the peak 
wavelength of the spectrum is at the OSA marker.
Change [
λ], the output wavelength of the TLS, to the wavelength of 
highest SSE (TLS@max_SSE_power) using the approximation: 
TLS@max_SSE_power = OSA@max_SSE_power + 0.5 nm]
10 Determine TLS@max_SSE_power as follows:
Set the Optical Spectrum Analyzer:
a Set the Sensitivity to -68 dBm.
b Set the resolution bandwidth to 0.5 nm.
c Set the center wavelength to OSA@max_SSE_power.
d Set the reference level to 0 dBm.
e Set the span to 6 nm.
f Record the spectrum.
11 Within the total spectrum, determine peak power, power@SSE_peak, 
and note the absolute value, |power@SSE_peak| in the Test Record.
Wavelength Specific These results are at the wavelength the TLS is 
set to for this measurement, and the OSA measurements, respectively.
NOTE