User`s guide
Cleaning Procedures for Lightwave Test and Measurement Equipment General Cleaning Procedure
260 Agilent 81600B Tunable Laser Source Family, Fourth Edition
How to clean instruments with a fixed 
connector interface
You should only clean instruments with a fixed connector interface when it 
is absolutely necessary. This is because it is difficult to remove any used 
alcohol or filaments from the input of the optical block.
It is important, therefore, to keep dust caps on the equip-ment at all times, 
except when your optical device is in use.
If you do discover filaments or particles, the only way to clean a fixed 
connector interface and the input of the optical block is to use compressed 
air.
If there are fluids or oil in the connector, please refer the instrument to the 
skilled personnel of the Agilent service team.
Only use clean, dry compressed air. Make sure that the air is free of 
dust, water, and oil. If the air that you use is not clean and dry, this 
can lead to filmy deposits or scratches on the surface of your 
connector interface. This will degrade the performance of your 
transmission system.
Never try to open the instrument and clean the optical block by 
yourself, because it is easy to scratch optical components, and 
cause them to become misaligned.
Both the surface and the jacket of the attached connector interface should 
be completely dry and clean.
CAUTION
NOTE