User`s guide
Definition of Terms Specifications
Agilent 81600B Tunable Laser Source Family, Fourth Edition 37
Absolute wavelength accuracy (continuous 
sweep mode)
The maximum difference between the “Logged wavelength” and the 
actual wavelength in “Continuous sweep mode”. Wavelength is defined 
as wavelength in vacuum.
Figure 7 Absolute wavelength accuracy (continuous sweep mode)
Conditions: As specified. No mode-hop.
Absolute wavelength accuracy (stepped 
mode)
The maximum difference between the displayed wavelength and the 
actual wavelength of the tunable lase source. Wavelength is defined as 
wavelength in vacuum.
Conditions: Constant power level. Other conditions as specified.
Measurement: Using a wavelength meter, averaging time
≥ 1s.
Attenuation
The nominal attenuation of the output power selected using the built-in 
attenuator.
The absolute wavelength accuracy of the low-SSE output
(if applicable) is the same as the absolute wavelength accuracy of the high 
power output (guaranteed by design).
NOTE
Applicable only to  Tunable Laser Source modules that include a built-
in attenator.
NOTE