User`s guide
Specifications Definition of Terms
46 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Conditions: Uninterrupted tunable laser source output power, constant power setting, 
constant temperature. Other conditions as specified.
Measurement: Using a wavelength meter, averaging time
≥ 1s.
Return loss
Specifies the ratio of the optical power incident to the tunable laser source 
output port at the wavelength set on the tunable laser source, to the power 
reflected from the tunable laser source output port.
Conditions: Tunable laser source output off. 
Side-mode suppression ratio
The ratio of optical power in the main mode to the optical power of the 
highest sidemode, expressed in dB:
Conditions: As specified.
Measurement: Using the Agilent Lightwave signal analyzer, by analyzing the heterodyning 
between the main signal and the highest sidemode within 0.1 GHz to 
6GHz.
Signal to source spontaneous emission 
(SSE) ratio
Specifies the ratio between signal power and maximum spontaneous 
emission (SSE) power. The SSE power is determined in a specified 
bandwidth within a ±3 nm window around the signal wavelength, where 
±1 nm around the signal wavelength are excluded, expressed in dB per 
specified bandwidth.
The relative wavelength accuracy of the low-SSE output
(if applicable) is the same as the relative wavelength accuracy of the high 
power output (guaranteed by design).
NOTE