User manual
38 G1369B LAN Interface User Manual
2Getting Started
Manual Configuration
With Telnet
Whenever a TCP/IP connection to the card is possible (TCP/IP parameters set 
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START 
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt: 
c:\>telnet <IP address> 
Figure 25 Telnet - Starting a session
where <IP address> may be the assigned address from a Bootp cycle, a 
configuration session with the Agilent Instant Pilot (G4208A) or Handheld 
Controller (G1323A/B), or the default IP address (see “Configuration 
Switches” on page 21).
When the connection was established successfully, the card responds with 
the following:
Figure 26 A connection to the module is made
3 To change a parameter follows the style:
parameter value
for example: ip 134.40.24.230