Technical information

2
Setting Up the LAN Interface
Step 3: Setting the Stored TCP/IP Address with a Telnet Session
Thermo Scientific Agilent 1200 Series Getting Connected Guide 9
Step 3: Setting the Stored TCP/IP Address with a Telnet Session
For an LC/MS system that includes Agilent 1200 Series LC modules and a Thermo Scientific
mass spectrometer, the IP address for the Agilent LAN interface must be set to 172.16.0.102.
Y To set the stored TCP/IP address for the LAN interface to 172.16.0.102
1. Set up the LAN interface initialization switches so that the LAN interface uses the default
IP address (see “Step 2: Setting Up the LAN Interface for a Telnet Sessionon page 7).
2. Ensure that the LAN interface is connected to the data system computer network card
labeled Surveyor MS.
3. Set the TCP/IP address for the data system computer network card to the following:
To set the TCP/IP address for the network card (for the Telnet session):
a. From the Windows taskbar, choose Start > Control Panel > Network Connection >
Local Area Connection.
The Local Area Connection dialog box appears.
b. On the General page, click Properties.
The Internet Protocol (TCP/IP) Properties dialog box appears.
c. Select the Use the following IP address option.
d. In the IP address box, type 192.168.254.100.
e. In the Subnet mask box, type 255.255.0.0.
f. Click OK to save the settings and close the Internet Protocol (TCP/IP) Properties
dialog box.
4. Initiate a Telnet session with the Agilent LAN interface as follows:
a. From the Windows taskbar, choose Start > Run.
The Run dialog box appears (see Figure 6).
IP address
192.168.254.100
Subnet mask
255.255.0.0
Note During this procedure, the MS detector temporarily loses communication with
the Xcalibur data system.