Installing the needle assembly
1
Push the silicon safety tube delivered with every needle 
over the needle.
2 Insert the loop capillary into the needle assembly and 
tighten the fitting hand tight.
3 Pinch the holder clamp and reinsert the needle assembly 
into the needle carrier.
4 Attach a 5/16 inch wrench to hold the position at the 
needle assembly. Use a 1/4 inch wrench to tighten the 
fitting of the loop capillary.