Agilent 1260 Infinity High Performance Autosampler User Manual 139
Maintenance
9
Installing the Metering Seal
3
Press the piston into the seal. 4 Reinstall the metering device / analytical head to the 
sampling unit by tightening the two fixing screws 
alternately with a 4 mm hex key. 
5 Connect the two capillaries to the metering device using 
a 1/4 inch wrench.
Next Steps:
6 Close the front door.
7 In the user interface exit the Change Metering device 
function and exit the maintenance mode. In the Lab 
Advisor software the Change Metering device function 
can be found in the Tools section.