User manual
188 Agilent 1260 Infinity High Performance Autosampler User Manual
12 Hardware Information
Early Maintenance Feedback
Early Maintenance Feedback
Maintenance requires the exchange of components which are subject to 
wear or stress. Ideally, the frequency at which components are exchanged 
should be based on the intensity of usage of the module and the analytical 
conditions, and not on a predefined time interval. The early maintenance 
feedback (EMF) feature monitors the usage of specific components in the 
instrument, and provides feedback when the user- selectable limits have 
been exceeded. The visual feedback in the user interface provides an 
indication that maintenance procedures should be scheduled.
EMF Counters
EMF counters increment with use and can be assigned a maximum limit 
which provides visual feedback in the user interface when the limit is 
exceeded. Some counters can be reset to zero after the required 
maintenance procedure.
Using the EMF Counters
The user- settable EMF limits for the EMF Counters enable the early 
maintenance feedback to be adapted to specific user requirements. The 
useful maintenance cycle is dependent on the requirements for use. 
Therefore, the definition of the maximum limits need to be determined 
based on the specific operating conditions of the instrument.
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two 
maintenance cycles. Initially the default EMF limits should be set. When 
instrument performance indicates maintenance is necessary, take note of 
the values displayed by the EMF counters. Enter these values (or values 
slightly less than the displayed values) as EMF limits, and then reset the 
EMF counters to zero. The next time the EMF counters exceed the new EMF 
limits, the EMF flag will be displayed, providing a reminder that 
maintenance needs to be scheduled.










