Agilent 1260 Infinity High Performance Autosampler User Manual 29
Site Requirements and Specifications
2
Specifications
Local Control Agilent Instant Pilot (G4208A) B.02.11 or above
Communications Controller-area network (CAN), RS-232C, 
APG Remote: ready, start, stop and 
shut-down signals, optional four external 
contact closures and BCD vial number 
out.
Safety and maintenance Extensive support for troubleshooting and 
maintenance is provided by the Instant 
Pilot, Agilent Lab Advisor, and the 
Chromatography Data System. 
Safety-related features are leak detection, 
safe leak handling, leak output signal for 
shutdown of pumping system, and low 
voltages in major maintenance areas.
GLP features Early maintenance feedback (EMF) for 
continuous tracking of instrument usage 
with user-settable limits and feedback 
messages. Electronic records of 
maintenance and errors.
Housing All materials recyclable.
Metering device Metering pump in high pressure flow path
1
Chromatographic conditions: Column: Agilent ZORBAX SB-C18, 2.1 x 50 mm1.8 µm 
(p/n 827700-902); mobile phase: A: 0.1 % TFA in water, B: 0.1 % TFA in acetonitrile; 
isocratic : %B=35 %; flow rate: 0.5 mL/min; temperature: 30 °C
2
UV-detection: Sample : 1200 ng/µL chlorhexidine (dissolved in mobile phase A), 1 µL injected and 
measured on G4212A DAD (10 mm cell); Wavelength: 257 nm +/- 4 nm; ref. 360 nm +/- 16 nm; slit 
4 nm, 10 Hz
3
MS-detection: Sample : 50 ng/µL chlorhexidine (dissolved in mobile phase A), 1 µL injected and 
measured on Agilent 6460 QQQ (in specified conditions); MRM 1: 505.5 → 170 (CE: 36 V); 
MRM 3: 505.5 → 201.2 (CE: 20 V); fragmentor: 150 V, delta EMV(+): 200 V
Tabl e 2 Performance Specifications (G1367E)
Type Specification Comment