User manual
Agilent 1260 Infinity High Performance Autosampler User Manual 67
Optimizing Performance
5
How to Configure the Optimum Delay Volume
How to Configure the Optimum Delay Volume
For very fast gradients over 0.5 min the delay volume of the system can 
be easily reduced without changing the physical configuration of the 
system. The change is achieved by changing the behavior of the 
autosampler.
The 270 µL delay volume of the autosampler is due to the flow path from 
the injection valve through the metering device, needle, needle seat and 
connecting capillaries back to the injection valve (see Figure 16 on 
page 68). To make an injection the valve switches from mainpass to 
bypass so that the metering device can draw the sample into the needle 
capillary. The injection is made when the valve switches back to mainpass 
and the sample is flushed onto the column. The valve remains in this 
position during analysis so that the autosampler is continually flushed and 
hence the gradient has to flow through this delay volume to reach the 
column. This can be eliminated by switching the injection valve from 
mainpass to bypass after the injection has been made and the injected 
sample has been flushed onto the column. In practice this can be done a 
few seconds after injection and is activated by selecting the “Automatic 
Delay Volume Reduction” (ADVR) function in the autosampler setup menu. 
The Flush- out Factor (typically 5 times injection volume) ensures that 
enough time is allowed to flush the sample out of the injector before 
switching to bypass. For instance a 1 µL injection under standard 
conditions effectively reduces the system delay volume by approximatly 
250 µL.










