User manual
72 Agilent 1260 Infinity High Performance Autosampler User Manual
5 Optimizing Performance
How to Achieve High Throughput
How to Achieve High Throughput
The injection can be optimized for speed remembering that drawing the 
sample too fast can reduce the reproducibility. Marginal gains are to be 
made here as the sample volumes used tend towards the smaller end of 
the range in any case. A significant portion of the injection time is the 
time taken with the needle movements to and from the vial and into the 
flush port. These manipulations can be performed while the previous 
separation is running. This is known as "overlapped injection" and it can 
be easily turned on from the autosampler setup screen in the control 
software. The autosampler can be told to switch the flow through the 
autosampler to bypass after the injection has been made and then after, 
for example, 3 minutes into a 4 minutes run to start the process of 
aspirating the next sample and preparing for injection. This can typically 
save 0.5 to 1 minute per injection.










