User manual
92 Agilent 1260 Infinity High Performance Autosampler User Manual
7 Error Information
General Error Messages
Compensation Sensor Short
Error ID: 0080
The ambient- compensation sensor (NTC) on the main board in the module 
has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the 
main board is dependent on ambient temperature. The change in 
resistance is used by the leak circuit to compensate for ambient 
temperature changes. If the resistance across the sensor falls below the 
lower limit, the error message is generated.
Fan Failed
Error ID: 0068
The cooling fan in the module has failed. 
The hall sensor on the fan shaft is used by the main board to monitor the 
fan speed. If the fan speed falls below a certain limit for a certain length 
of time, the error message is generated.
Depending on the module, assemblies (e.g. the lamp in the detector) are 
turned off to assure that the module does not overheat inside.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service 
representative.
Probable cause Suggested actions
1 Fan cable disconnected.
Please contact your Agilent service 
representative.
2 Defective fan.
Please contact your Agilent service 
representative.
3 Defective main board.
Please contact your Agilent service 
representative.