User's Manual

122 Chapter 1
GPIB Command Dictionary
CDANalyzer subsystem
:GAIN <string>
:GAIN?
These commands set/query the gain of the code domain analyzer. This
command will only be enabled when autoranging is turned off
(CDAN:GAIN:MODE ‘Hold’).
This command may conflict with the CDMA analyzers settings
CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging
is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code
Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the
‘Auto’ setting. When autoranging is turned off, setting either analyzers
gain will automatically change the others to the same setting.
Use the CAN:POW:GAIN command when making an average power
measurement.
Syntax
CDAN:GAIN ‘0 dB’
‘6 dB’
‘12 dB’
‘18 dB’
‘24 dB’
‘30 dB’
‘36 dB’
CDAN:GAIN?
Screen/field equivalent
CDAN:GAIN sets the lower subfield of the Gain field of the CODE DOM
screen, when the CDMA standard is set to “IS-95 Only”. This subfield is
only selectable when the upper subfield is set to Hold.