User's Manual

Chapter 1 133
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:GAIN[:SETTing] <string>
:IS2000:GAIN[:SETTing]?
For IS-2000 only. These commands set/query the gain of the Code
Domain analyzer. This command will only be enabled when
autoranging is turned off (CDAN:IS2000:GAIN:MODE ‘Hold’).
This command may conflict with the CDMA analyzers settings
CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging
is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code
Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the
‘Auto’ setting. When autoranging is turned off, setting either analyzers
gain will automatically change the others to the same setting.
Use the CAN:POW:GAIN command when making an average power
measurement.
Syntax
CDAN:IS2000:GAIN[:SETT] ‘0 dB’
‘6 dB’
‘12 dB’
‘18 dB’
‘24 dB’
‘30 dB’
‘36 dB’
CDAN:IS2000:GAIN[:SETT]?
Screen/field equivalent
CDAN:IS2000:GAIN[:SETT] sets the lower subfield of the Gain image
of the CODE DOMAIN screen. This subfield can only be set when the
upper subfield is set to Hold when the CDMA standard is set to
“IS-2000”.