User's Manual

Chapter 1 Getting Started
Operating the Counter
1-18 Operating Guide
1
NOTES The Counter displays CH2 NO SIGNAL or CH1 NO SIGNAL and shuts
down all unnecessary circuits when a signal of insufficient amplitude
(or no signal) is applied to the corresponding input. This extends the
reliability of the affected components, and if the Battery option is
installed, extends the length of time the Counter can operate from
the batteries.
When the frequency of a signal applied to the Channel 2 input exceeds the
maximum rated frequency for the instrument, the Counter displays CH2
TOO HIGH.
CAUTION The 2.9 mm Planar Crown* connector used for the Channel 2 input on the
Agilent 53152A must be handled with care to prevent damage and/or
contamination, especially since it acts as a wave guide as well as an
electrical connection. Observe the following precautions when handling
this connector:
1. If you remove the outer portion of the connector, do not touch the
exposed surfaces of either part of the connector with your bare skin or
any material that is not intended for cleaning this type of connector.
2. Avoid dropping or striking either portion of the connector.
If the connector becomes contaminated, it can be cleaned with isopropyl
alcohol and a lint-free cloth or other suitable cleaning implement.
* Planar Crown
®
is a registered trademark of Weinschel Corp.