User's Manual
2 - Verification and Performance Tests
20
Performance Test Record Forms
Model Agilent 66332A OR
Agilent6632B
Report No _______________ Date __________________
Test Description Minimum
Specs.
 Results* Maximum
Specs.
Measurement
Uncertainty
 Constant Voltage Tests
Voltage Programming and Readback
 Low Voltage (0V) Vout
 Front Panel Display Readback
 High Voltage (Full Scale) Vout
 Front Panel Display Readback
− 10 mV
Vout − 3 mV
19.980 V
Vout − 9 mV
__________
__________
__________
__________
+ 10 mV
Vout + 3 mV
20.020 V
Vout + 9 mV
1.6 µV
1.6 µV
335 µV
335 µV
Load Effect
− 2 mV
__________ + 2 mV
20 µV
Source Effect
− 0.5 mV
__________  + 0.5 mV
20 µV
PARD (Ripple and Noise)
 Peak-to-Peak
 RMS
0 mV
0 mV
__________
__________
+ 3 mV
+ 0.3 mV
872 µV
50 µV
Transient Response
 Voltage in 100 µs
0 mV __________ + 20 mV 3 mV
 Constant Current Tests
Current Programming and Readback
 Low current (0A) Iout
 Readback Accuracy @ Iout
 High Current (Full Scale) Iout
 Readback Accuracy @ Iout
− 2.0 mA
Iout − 0.5 mA
4.9955 A
Iout − 10.5 mA
__________
__________
__________
__________
+ 2.0 mA
Iout + 0.5 mA
5.0045 A
Iout + 10.5 mA
15.2 µA
15.2 µA
818.7 µA
818.7 µA
Current Sink Readback
Isink − 11.1mA
__________
Isink + 11.1mA
818.7 µA
20 mA Range Current Readback
 Readback Accuracy @ 0 A
 Readback Accuracy @ + 20 mA
 Readback Accuracy @ − 20 mA
− 2.5 µA
Iout − 22.5 µA
Iout − 22.5 µA
__________
__________
__________
+ 2.5 µA
Iout + 22.5 µA
Iout + 22.5 µA
0.1 µA
1.7 µA
1.7 µA
PARD (Current Ripple and Noise)
 RMS
0 mA __________ + 2.0 mA
250 µA
Load Effect
− 1.0 mA
__________ + 1.0 mA
2.5 µA
Source Effect
 − 0.5 mA
__________
+ 0.5 mA
2.5 µA
* Enter your test results in this column