Troubleshooting guide
Lab Exercise: Advanced Reporting 
Automated Spectral Library Search 
131 
to the calibration table for an identically named entry. If one is found, the data are 
used to calculate and report the amount. 
6)  OK this dialog box. Open the library you intend to use in the report and 
check that the names of the calibrated peaks are identical with the library 
names. In other words, check your calibration table as well as the entries in 
the library. 
7)  Under Specify Report, select Library Search. OK the dialog box. Resave 
your method. Print the report. 
8)  Now, set up a method where the search mode is Target compound analysis 
using a calibration table. 
This search mode is only for those compounds in the calibration table. Peaks are 
only reported as identified if confirmed by spectral comparison. 
9)  Print the report. 










