Troubleshooting guide
Lab Exercise: Integration 
Auto Integration 
77 
5)  Using the Events Table drop box, select signal DAD1 B Specific. 
6)  To Auto Integrate, select the Integration tool then the Auto Integrate tool   
or select the Integration menu, then Auto Integrate. When asked if you 
want to save the changed events table, reply No. 
7)  View the events and chromatogram that were created for Signal B. 
8)  Now see what happens when you add the recommended For All Signals 
settings: 
Tangent Skim Mode: New Exponential 
Tail Peak Skim Height Ratio: 3.00 
Front Peak Skim Height Ratio: 6.00 
Skim Valley Ratio: 20.00 
Baseline Correction: Advanced 
Peak to Valley Ratio 500.00 
Note: Auto Integrate does not set the For All Signals Integration Events. You 
must set these manually. 
10)  Use Auto Integrate again. Again, do not save the changed events table. 
Print the chromatogram to notice any differences. You may want to print 
this chromatogram for comparison and record the event settings below. 
Did you see any improvement? 
Initial Slope Sensitivity__________ 
Initial Peak Width __________ 
Initial Area Reject __________ 
Initial Height Reject __________ 
Initial Shoulders __________ 