Technical data
Agilent B1500A/B1505A Programming Guide, Edition 11 4-33
Command Reference
AAD
AAD
This command is used to specify the type of the A/D converter (ADC) for each 
measurement channel.
Execution 
Conditions
Enter the AIT command to set up the ADC.
Syntax  AAD chnum,type
Parameters  chnum :  SMU measurement channel number. Integer expression. 1 to 10 or 101 
to 1001. See Table 4-1 on page 4-16.
type :  Type of the A/D converter. Integer expression. 0, 1, or 2.
0: High-speed ADC for high speed DC measurement. Initial setting.
1: High-resolution ADC. For high accurate DC measurement. 
Not available for the HCSMU, HVSMU, and MCSMU.
2: High-speed ADC for pulsed-measurement
The pulsed-measurement ADC is automatically used for the pulsed spot, pulsed 
sweep, multi channel pulsed spot, multi channel pulsed sweep, or staircase sweep 
with pulsed bias measurement, even if the AAD chnum,2 command is not executed.
The pulsed-measurement ADC is never used for the DC measurement. Even if the 
AAD chnum,2 command is executed, the previous setting is still effective.
Example 
Statements 
OUTPUT @B1500;"AAD 1,0"
OUTPUT @B1500;"AAD 1,1"
AB 
The AB command aborts the present operation and subsequent command execution.
This command stops the operation now in progress, such as the measurement 
execution, source setup changing, and so on. But this command does not change the 
present condition. For example, if the B1500 just keeps to force the DC bias, the AB 
command does not stop the DC bias output.
Syntax  AB
Example 
Statements 
OUTPUT @B1500;"AB"