DSMPLFLUSH
Parameters  chnum :  SMU channel number. Integer expression. 1 to 10. See Table 4-1 on 
page 4-16.
type :  Event type. Integer expression.
1: Start of pulse output
count :  Number of events. Integer expression. 0 disables the signal monitor.
Example 
Statements 
If the following command is executed, the channel 5 starts the signal monitor when 
the 10th pulse output starts.
OUTPUT @B1500;"DSMPLARM 5,1,10"
Remarks The signal monitor operation is performed as a part of the MM27/MM28 
measurement.
If the signal monitor operation overlaps a MM27/MM28 measurement period, the 
overlapped monitor data is averaged and the result is set as a MM27/MM28 
measurement data. If there is no overlap, a dummy data is set.
The MM27/MM28 measurement data is directly sent to the data output buffer 
regardless of the signal monitor operation. 
The channel can keep all data obtained by one execution of the signal monitor. And 
the data is sent to the data buffer by the DSMPLFLUSH command.
DSMPLFLUSH
Available for HVSMU, HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and 
UHVU.
The DSMPLFLUSH command sends the signal monitor data to the data output 
buffer. 
Execution 
Conditions 
The DSMPLARM command must be executed before this command.
Syntax  DSMPLFLUSH chnum
Parameters  chnum :  SMU channel number. Integer expression. 1 to 10. See Table 4-1 on 
page 4-16.
Response All data obtained by one execution of the signal monitor is returned. One data 
contains the compliance side data and the force side data in this order. Time stamp 
data is not included. The data output format depends on the FMT command setting.