Technical data
4-76 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
DSMPLSETUP
If there is no monitor data, a dummy data is returned.
Example 
Statements 
OUTPUT @B1500;"DSMPLFLUSH 5"
DSMPLSETUP 
Available for HVSMU, HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and 
UHVU.
The DSMPLSETUP command sets the signal monitor function for the specified 
channel. This function is used to monitor the channel input/output during the 
measurement specified by the MM command. By using the DSMPLARM and 
DSMPLFLUSH commands, the channel measures its voltage and current, and 
repeats it the specified count in the specified interval. Then the voltage and current 
measurements are performed at the same time.
Syntax  DSMPLSETUP chnum,count,interval[,delay]
Parameters  chnum :  SMU channel number. Integer expression. 1 to 10. See Table 4-1 on 
page 4-16.
count :  Number of sampling. Integer expression. 1 to 4000 for HVSMU. 
1 to 2000 for HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and 
UHVU.
interval :  Sampling interval. Numeric expression. 6 μs to 12 μs in 2 μs resolution 
for HVSMU. 2 μs to 12 μs in 2 μs resolution for HCSMU, MCSMU, 
DHCSMU, UHCU, HVMCU, and UHVU.
delay :  Delay time. Numeric expression. - maximum count value × interval to 
+ maximum count value × interval, in 2 μs resolution. Default setting is 
0.
Set 110 μs to start the signal monitor at the timing to start the transition 
from the pulse base value to the pulse peak value.
Example 
Statements 
OUTPUT @B1500;"DSMPLSETUP 5,100,6E-6,1E-4"
DV 
The DV command forces DC voltage from the specified SMU.
Execution 
Conditions 
The CN/CNX command has been executed for the specified channel.