Technical data
Agilent B1500A/B1505A Programming Guide, Edition 11 4-77
Command Reference
DZ
If the output voltage is greater than the allowable voltage for the interlock open 
condition, the interlock circuit must be shorted.
Syntax  DV chnum,vrange,voltage[,Icomp[,comp_polarity[,irange]]
Parameters  chnum :  SMU source channel number. Integer expression. 1 to 10 or 101 to 
1001. See Table 4-1 on page 4-16.
vrange :  Ranging type for voltage output. Integer expression. The output range 
will be set to the minimum range that covers voltage value. For the 
limited auto ranging, the instrument never uses the range less than the 
specified range. See Table 4-4 on page 4-21.
voltage :  Output voltage value (in V). Numeric expression. See Table 4-7 on 
page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table 
4-15 on page 4-28 for each measurement resource type.
Icomp :  Current compliance value (in A). Numeric expression. See Table 4-7 on 
page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table 
4-15 on page 4-28 for each measurement resource type. If you do not 
set Icomp, the previous value is used. 0 A is not allowed for Icomp.
comp_ 
polarity :  Polarity of current compliance. Integer expression.
0:  Auto mode (default setting). The compliance polarity is 
automatically set to the same polarity as voltage, regardless of the 
specified Icomp. If voltage=0 V, the polarity is set to positive.
1:  Manual mode. Uses the polarity of Icomp you specified.
irange :  Current compliance ranging type. Integer expression. The compliance 
range will be set to the minimum range that covers Icomp value. For the 
limited auto ranging, the instrument never uses the range less than the 
specified range. See Table 4-5 on page 4-22.
Example 
Statements 
OUTPUT @B1500;"DV 1,0,20,1E-6,0,15"
OUTPUT @B1500;"DV 2,12,10"
DZ 
The DZ command stores the settings (V/I output values, V/I output ranges, V/I 
compliance values, and so on) of the specified channels, and sets the channels to 0 
V. The settings can be recovered by using the RZ command. The stored settings are 
cleared by using a device clear (HP BASIC CLEAR) command, *RST, RZ, CL, CA, 
or *TST?.










