Technical data
4-86 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERHVCA?
ERHVCA?
This query command returns the slot numbers for the modules connected to the V 
Control, I Control, and HVSMU inputs of Agilent N1266A HVSMU current 
expander.
Syntax  ERHVCA?
Query Response  vsmu,ismu,hvsmu<CR/LF^EOI>
vsmu :  Slot number for the MC/HCSMU connected to the V Control input.
ismu :  Slot number for the MC/HCSMU connected to the I Control input.
hvsmu :  Slot number for the HVSMU connected to the HVSMU input.
Example 
Statements 
OUTPUT @B1500;"ERHVCA?"  
ENTER @B1500;vsmu,ismu,hvsmu
ERHVCTST?
This query command performs the self-test of Agilent N1266A HVSMU current 
expander, and returns the execution results. After this command, read the result 
soon. 
Execution 
Conditions
The ERHVCA command has been executed before this command.
Syntax  ERHVCTST?
Query Response  result<CR/LF^EOI>
0: Pass
1: Fail
Example 
Statements 
OUTPUT @B1500;"ERHVCTST?"
ENTER @B1500;result
ERHVP
This command sets the input-to-output connection path of Agilent N1266A 
HVSMU current expander.
Execution 
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage 
for the interlock open condition, or voltage compliance set to more than it). 