Error Messages
Operation Error
273  Search and sync output modes must be the same.
The primary search source channel and the synchronous source channel 
must be different, and they must be set to the same output mode 
(voltage or current).
274  Search sync source is overflow.
Set the search sources so that the same output range is set to both 
primary and synchronous search sources.
275  Search target must be compliance value or less.
The search target value must be less than or equal to the compliance 
value of the search monitor channel. Correct the search target value or 
the compliance value.
276  Start and stop must be different.
Set different values for the search start and stop values.
277  Step must be output resolution or more.
Set the search step value to the output resolution or more.
278  Search and sync channels must be different.
Set the search source and the synchronous source to different channels.
279  Search monitor mode must be compliance side.
Send the LGI/BGI command to set the voltage source search monitor 
channel, or send the LGV/BGV command to set the current source 
search monitor channel.
280  Send WDCV to set CV sweep source.
Before triggering the multi frequency CV measurement, send the 
WDCV command to set the DC voltage sweep source.
281  Send PDCV to set CV pulse source.
Before triggering the pulsed spot C measurement, send the PDCV 
command to set the pulsed voltage source.
282  Send PWDCV to set CV pulse sweep source.
Before triggering the pulsed CV measurement, send the PWDCV 
command to set the pulsed voltage sweep source.