Technical data
5-26 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
5309  CRC for N1265A EEPROM Vm correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies 
service center.
5310  CRC for N1265A EEPROM Im correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies 
service center.
5311  Specified module is already assigned to voltage control or current 
control of HVMCU.
Specify a free MCSMU/HCSMU.
5312  Different module must be assigned to voltage control and current 
control of HVMCU.
Specify a free MCSMU/HCSMU.
5313  Do not execute CN/CNX to the current control module set to 
ERHVCA.
Execute the CN/CNX command to the module that is set to voltage 
control by the ERHVCA command.
5314  HVSMU Current Expander output must be open state.
Open the input-output path before switching the HVSMU series 
resistor.
5315  Execute CL to the module set to ERHVCA.
Execute the CL command to the modules that will be specified by the 
ERHVCA command.
5316  CRC for N1266A EEPROM Vm correction segment failed.
N1266A might be defective. Contact your nearest Agilent Technologies 
service center.
5317  CRC for N1266A EEPROM Im correction segment failed.
N1266A might be defective. Contact your nearest Agilent Technologies 
service center.
5318  Over voltage is detected in Selector Output High Force, Low Force, or 
Low Sense terminal of N1265A.
Remove the causes of overvoltage.