Remote Mode Functions
Measurement Modes
NOTE The WT command sets the hold time, delay time, and step delay time. The WM 
command sets the automatic abort function and the output after measurement. The 
WV command sets the sweep voltage, and the WI command sets the sweep current. 
The start and stop values must have the same polarity for log sweep.
To Use 
Synchronous 
Sweep Source
One more channel can be set up as a staircase sweep source that has the output 
synchronized with the staircase sweep. Refer to “Synchronous Output” on page 
2-44. After the measurement, the synchronous sweep source forces the start or stop 
value, as same as the primary sweep source, and keeps it.
Figure 2-5 Synchronous Sweep
NOTE The WSV command sets the sweep voltage, and the WSI command sets the sweep 
current. You can use the same output mode (voltage or current) as the primary 
sweep. The start and stop values must have the same polarity for log sweep.
To Stop Sweep 
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on 
page 2-46.
Even if the automatic abort function is disabled, the B1500 automatically stops 
measurement if power compliance is enabled for the sweep source and the power 
compliance or an automatic abort condition is detected.
Voltage or current
Start value
Time
Stop value
Primary sweep
Synchronous sweep
Previous value
Trigger (e.g. XE)
WT/WM/WV/WI
WSV/WSI
Voltage or current
Start value
Stop value